The YL9130 Column Compartment with Peltier cooling provides a wide range of temperature settings from 4°C up to 90°C. Effective temperature control ensures retention time reproducibility and reliable data. Column temperature can also be used to improve separation performance and analysis times.
Specifications Operating principle: Peltier elements with 200mm heat exchanger Temperature range: 4°C (Cooling) - 90 °C Temperature accuracy: ±0.5°C Temperature stability: ±0.05°C Temperature programs: 40 Steps Column capacity: 2 x 250mm, max 3 x 150mm Column switching: Maximum two automatically 6-port valves (optional) Heat-up time: 16 minutes from 4°C to 90°C Cool-down time: 13 minutes from 90°C to 4°C Communication: LAN Safety & maintenance: Leak detection, Diagnostics, Error detection Dimensions (W x H x D): 385mm x 160mm x 565mm Line voltage: 95 - 240 Volt AC ± 10%, automatic voltage selection Line frequency: 50/60Hz, ±5% Power consumption: 150W